Structured Illumination Microscopy (SIM)

NewsGuard 100/100 Score

Structured Illumination Microscopy (SIM) is a tool which breaks the diffraction resolution barrier to provide resolutions of up to 20nm.

Structured illumination microscopy (SIM) images of a bovine pulmonary artery endothelial cell stained with fluorescent dyes for mitochondria, actin, nucleus.Image Credit: Micha Weber / Shutterstock

The resolution of conventional microscopes is limited by the diffraction of light. This resolution is 200 nm laterally and 500 nm in the axial direction. Thus, it is difficult to visualise any object which lies below the range of this resolution.

The size of most of the structures inside a cell is below this limit, which makes light microscopy an inappropriate tool to visualize sub-cellular structures.

Principles of SIM

Structured Illumination Microscopy or SIM is based on wide-field microscopy where a movable diffraction grating is placed in the path of the laser beam. Then, the zeroth order and first order rays from the diffraction grating are allowed to fall on the sample.

The light then undergoes interference (light waves superimpose on each other leading to a light of greater or lesser wavelength) with each other at the sample plane. This pattern of light created by the superimposition of light on the sample is called ‘Moire effect’.

Several such raw images are taken by changing the angle or orientation of this ‘structured light’. This is done by changing the angle of the movable diffraction grating. Several such images taken at different angles are then reconstructed post-imaging to generate a super-resolved image. The high frequency image extracted from the raw data can have almost double the resolution of the conventional diffraction limited microscopes.

Advantages of SIM

Improved resolution

SIM provides a twice improved scale of resolution. In two dimensional SIM, resolution in the lateral axes is improved, while in 3D SIM, both lateral and axial resolution show improvement. The resolution in SIM is 100 and 250 nm in the lateral and axial directions.

Fast rate of imaging

The rate of imaging is very fast for both 3D imaging and 4D imaging. The fourth dimension refers to time, or time-lapse imaging. This factor is very crucial for imaging live samples where biological processes may need to be captured at high speeds.

Conventional fluorophores can be used

Fluorophores which are used for conventional microscopy can be used for imaging in SIM. This is an added advantage as no additional resources have to be employed for preparing samples. This is different from some of the other super-resolution microscopy techniques, where specialized fluorophores have to be generated.

Simultaneous use of multiple fluorophores

In SIM, three fluorophores can be imaged simultaneously. This means that three intra-cellular structures can be tracked together using SIM, while this is usually limited to one or two fluorophores in case of other super-resolution tools.

Live imaging

It is possible to image live samples and record different cellular processes. This feature is not available in some of the super-resolution techniques, such as STED.

Disadvantages of SIM

Reduced resolution compared to other super-resolution tools

The lateral resolution achieved by other super-resolution techniques, such as STORM/PALM and STED is almost 20 nm which is higher than SIM.

Imaging artifacts can arise

The super-resolution is achieved not optically, but through image processing after the imaging. Thus, artifacts could be generated during the image reconstruction. Also, errors in grating, calibration of the system, mismatch of the refractive index, or poor quality of the sample can also generate errors in the final image.

Sensitive to out-of-focus light

It is very sensitive to out of focus light thus the signal is not very clear if the sample is densely populated with the molecules being imaged.

Longer processing times

SIM requires a processing time to 1 to 30 seconds when the image reconstruction can be applied to finally generate the final image.

Further Reading

Last Updated: Feb 26, 2019

Dr. Surat P

Written by

Dr. Surat P

Dr. Surat graduated with a Ph.D. in Cell Biology and Mechanobiology from the Tata Institute of Fundamental Research (Mumbai, India) in 2016. Prior to her Ph.D., Surat studied for a Bachelor of Science (B.Sc.) degree in Zoology, during which she was the recipient of an Indian Academy of Sciences Summer Fellowship to study the proteins involved in AIDs. She produces feature articles on a wide range of topics, such as medical ethics, data manipulation, pseudoscience and superstition, education, and human evolution. She is passionate about science communication and writes articles covering all areas of the life sciences.  

Citations

Please use one of the following formats to cite this article in your essay, paper or report:

  • APA

    P, Surat. (2019, February 26). Structured Illumination Microscopy (SIM). News-Medical. Retrieved on May 05, 2024 from https://www.news-medical.net/life-sciences/Structured-Illumination-Microscopy-(SIM).aspx.

  • MLA

    P, Surat. "Structured Illumination Microscopy (SIM)". News-Medical. 05 May 2024. <https://www.news-medical.net/life-sciences/Structured-Illumination-Microscopy-(SIM).aspx>.

  • Chicago

    P, Surat. "Structured Illumination Microscopy (SIM)". News-Medical. https://www.news-medical.net/life-sciences/Structured-Illumination-Microscopy-(SIM).aspx. (accessed May 05, 2024).

  • Harvard

    P, Surat. 2019. Structured Illumination Microscopy (SIM). News-Medical, viewed 05 May 2024, https://www.news-medical.net/life-sciences/Structured-Illumination-Microscopy-(SIM).aspx.

Comments

The opinions expressed here are the views of the writer and do not necessarily reflect the views and opinions of News Medical.
Post a new comment
Post

While we only use edited and approved content for Azthena answers, it may on occasions provide incorrect responses. Please confirm any data provided with the related suppliers or authors. We do not provide medical advice, if you search for medical information you must always consult a medical professional before acting on any information provided.

Your questions, but not your email details will be shared with OpenAI and retained for 30 days in accordance with their privacy principles.

Please do not ask questions that use sensitive or confidential information.

Read the full Terms & Conditions.