Helping detect gastrointestinal disorders, the GastroCH4ECK® is the newest member of the Gastrolyzer® family. It is a portable breath analysis monitor that measures accurate and real-time combined CH4, H2 and O2. This monitor is intended for multi-patient use by healthcare professionals in a clinical environment.
Download the Brochure for More Information
Why is it important?
The monitor provides accurate and real-time combined CH4, H2 and O2 monitoring.
The GastroCH4ECK® is a portable breath CH4, H2 and O2 monitor that helps health professionals to accurately detect a range of gastrointestinal disorders. An O2 reading is taken to motivate patients into providing an end-tidal sample. If they are unable to do this the GastroCH4ECK® will automatically correct the reading saving the patient from embarrassment and having to carry out another test.
A percentage of patients do not produce hydrogen and therefore would not be diagnosed by the use of hydrogen breath testing (HBT) alone. Therefore the role of methane is to ensure that patients who are non-hydrogen producers, but produce methane instead, are not miss-diagnosed.
Most monitors measure hydrogen alone, which identifies less than 50% of sufferers. Methane is detected in 30%-50% of the healthy adult population worldwide.
Why GastroCH4ECK®?
The GastroCH4ECK® can be used to help detect the following disorders:
- Lactose intolerance
- Fructose intolerance
- Glucose intolerance
- Irritable Bowel Syndrome (IBS)
- Small Intestinal Bacterial Overgrowth (SIBO)
Benefits for the user
- The monitor provides hydrogen and methane breath sampling with instant results.
- The breath bag sampling options allows for simultaneous testing of larger groups or remote testing for patients who cannot provide samples directly in person.
- The monitor only requires calibration once per month, saving precious time and calibration gas costs.
- The monitor has a fast warm up time of less than 30 minutes.
- The monitor provides O2 correction for accurate results.
- There is a real-time traffic light dial and smiley face system to motivate patients and indicate how the test is going.
- The consumables contain anti-bacterial filters for optimum infection control.
Download the Brochure for More Information
Technical specification
Concentration range |
CH4 0-200 pm; H2 0-200 ppm; O2 14-23% |
Power |
230 V/100 V, 50 Hz-60 Hz, 0.5-1.0 A |
Fuse |
T 3.15 AH |
Start-up time |
≤2 minutes |
Calibration frequency |
Every 4 weeks |
Display |
Full colour touchscreen |
Detection principle |
Electrochemical sensor (H2 & O2); Laser (CH4) |
Accuracy: CH4 |
Resolution: 1 ppm; Accuracy: ±10% of reading; Repeatability: <5% difference on consecutive readings |
Accuracy: H2 |
Resolution: 1ppm; Accuracy: ±10% of reading; Repeatability: <5% difference on consecutive readings |
Accuracy: O2 |
Resolution: 0.1%; Accuracy: ±10% of reading; Repeatability: <5% difference on consecutive readings |
CO cross-sensitivity (H2 only) |
<4% |
Temperature range |
Operating: 15-35 °C (59-95 °F); Storage: 0-40 °C (32-104 °F) |
Pressure range |
Operating: 912-1114 mbar (Atmospheric ±10%); Storage: 912-1114 mbar (Atmospheric ±10%) |
Sensor operating life |
CH4 5 years; H2 2 years; O2 2 years |
Dimensions |
Approx. 474 x 310 x 135 mm |
Weight |
Approx. 8.5 kg |
Materials |
Case: aluminium |
Classification |
Class I ME equipment: (externally powered) Type BF applied part; Method of sterilisation (not suitable for sterilisation); Not suitable for use in an oxygen rich environment; Intended for continuous use |
Warm-up time |
<45 seconds |
Warranty |
GastroCH4ECK® without sensors: 2 years; CH4, H2 & O2 sensors: 1 year |