Clarinet is a clinical middle ear analyzer, featured with all the tests needed for a deep and accurate investigation of middle ear disorders. Automatic and manual tympanometry, acoustic reflex threshold, decay and latency examinations, ETF for intact and perforated eardrums, high frequency and multi-component tympanometry are all available and full results can be observed on a wide color touch screen display.
THE DISPLAY
Clarinet incorporates a large high-resolution color graphic display, giving a clear and detailed picture of the exam in progress. And with touch screen input, interaction with the instrument is extremely simple.
THE PROBE
Clarinet features an extremely light, small and sturdy metal probe, which can be very rapidly disassembled and cleaned. The probe is connected to the small Control Box, through which you can select the ear and start the exam.
ENDLESS FLOW PUMP
Clarinet immediately impresses for the incredible speed of its pump: the sweep speed can reach 600 daPa / sec, without compromising on accuracy and precision of tympanogram plots. The pump features an endless airflow technology, improving the execution of the test when the seal is not optimal.
GO IN-DEPTH WITH INVESTIGATIONS
High-frequency and multi-component tympanometry, evaluation of multiple stimuli averaged reflex latencies together with the up-to-date measure of reflex plot angles of incidence are just some of the Clarinet advanced features that will take your clinical investigation to new levels.
NOAH & DAISY INTEGRATION
Simply connect Clarinet to a USB port on your computer (no driver needed), and you can transfer exam results into your Daisy or Noah database. And with the live view feature, you can have a wide view of the examination in progress on the computer display.
TECHNICAL DATA
PROBE TONE
|
226 HZ AT 85 DB SPL
678HZ, 800HZ, 1000HZ (PLUS)
|
COMPL. RANGE
|
0.2 TO 8.0 ML / 0.9 TO 15.0 MMHO
|
PRESSURE RANGE
|
FROM -600 TO +400 DAPA
|
PUMP CONTROL
|
AUTOMATIC AND MANUAL
|
PRESS. CHANGE RATES
|
15, 50, 100, 200, 300, 400, 600 DAPA/SEC, AUTO (200-600)
|
PRESS. SWEEP
|
BIDIRECTIONAL
|
TYMPANOMETRY MEASURES
|
ADMITTANCE [Y]
SUSCEPTANCE [B] (PLUS)
CONDUCTANCE [G] (PLUS)
|
ACOUSTIC REFLEX TESTS
|
MANUAL TEST
AUTOTHRESHOLD
FIXED INTENSITY
GROWING INTENSITY
REFLEX DECAY
REFLEX LATENCY
|
STIMULATION
|
IPSI AND CONTRALATERAL
|
STIMULI
|
250 (CONTRA), 500, 1K, 2K, 4K, 6K, 8K (CONTRA) HZ, BBN, HPN, LPN
|
MAXIMUM INT
|
110 DB HL (IPSI); 120 DB HL (CONTRA)
|
STIM. DURATION
|
FROM 0.5 TO 2.5 SEC (SELECTABLE)
REFLEX DECAY: 10 OR 20 SEC
|
AUTOMATIC TESTS
|
QUICK A AND B (SEQ. OF TYMPAN. AND REFLEX)
|
ETF TEST
|
INTACT OR PERFORATED EARDRUMS
|
MEMORY
|
3 TYMPAN. + 16 REFLEX + 4 DECAY + 4 LATENCY RECORD. PER EAR
|
DISPLAY TYPE
|
COLOR GRAPHICAL WITH TOUCH SCREEN
|
DISPLAY SIZE, RESOLUTION
|
7”, 800 X 480
|
PRINTER
|
BUILT-IN THERMAL PRINTER (OPTIONAL)
|
LANGUAGE
|
MULTILINGUAL INTERFACE
|
PC INTERFACE
|
USB (DRIVERLESS)
|
COMPATIBLE SOFTWARE
|
INVENTIS DAISY OR NOAH WITH DEDICATED MODULE
|
DIMENSIONS
|
32 X 32 X 15 CM - 12.6 X 12.6 X 5.9 IN
|
WEIGHT
|
2.2 KG / 4.9 LBS
|
STANDARDS
|
TYMPAN.: EN 60645-5 / ANSI S3.39, TYPE 1
SAFETY: EN 60601-1
EMC: EN 60601-1-2
|
CLASSIFICATION
|
CLASS IIA (MDD 93/42)
|