- Provides test results within seconds
- Allows testing from infants to adults
- Customizable to suit individual needs
- Full 10.4'' touch-based interface for an intuitive handling
- Innovative probe lights allow patient-focused operation
- Probe tone 226 Hz, optional high frequency probe tone 1 kHz
- Acoustic reflex frequencies 500 Hz, 1 kHz, 2 kHz, 4 kHz (ipsi- and optional contralateral)
- High resolution and quick screen transitions
- Built-in printer
TECHNICAL SPECIFICATIONS
Tympanometry |
Probe Frequency and Intensity |
226 Hz ± 1 %, 85 dBSPL (69 dBHL) ± 1.5 dB
|
Optional High Frequency |
1 kHz ± 1%, 69 dBSPL (69 dBHL) ± 1.5 dB
|
Pressure Range |
-400 to +200 daPa |
Accuracy of Measurement |
± 5 % or 0.1 ml |
Volume Range |
0.1 to 8.0 ml |
Tymp Compliance Range |
0.1 to 8.0 ml at 226 Hz 0.1 to 15.0 mmho at 1000 Hz |
Test Time |
3 - 5 seconds |
Acoustic Reflex |
Test Frequencies |
0.5, 1, 2, 4 kHz ± 1 % |
Test Method |
Ipsilateral; optional contralateral |
Test Level Ipsilateral |
70 to 105 dBHL |
Test Level Contralateral |
70 up to 120 dBHL with contra phone receiver |
Intensity Setting |
Automatic or fixed |
Ipsilateral Reflex Test |
With AGC |
General |
Display |
10.4" Graphic LED-Display with resistive touchscreen |
PC Interface |
USB |
Printer |
Fast, virtually silent 4 inch thermal printer |
Standards |
IEC 60601-1, class I, type B, IEC 60645-5, type 2, according to medical device directive 93/42/EEC
|
Device |
Probe |
Lightweight hand-held probe with light indicators |
Power Supply |
Mains 100 to 240 V~ ± 10 %, 50-60 Hz ± 10 % |
Dimensions Device |
W 30 x D 34.5 x H 14.8 cm |
Weight Device |
3.2 kg |
Accessories |
Standard |
Pen Probe, built-in calibration cavity, Sanibel ear tip kit, built-in printer and roll of thermal paper, mains cable, probe floss kit, MAICO USB flash drive kit, Impedance Module software, USB-cable, cleaning cloth, touchpen
|
Optional |
Contralateral phones: DD45 C, CIR22, IP30
|
MAICO touchTymp Intro
touchTymp MI 24 | MAICO Products