Infrared (IR) spectroscopy is a widely used analytical technique both for identification and quantification of a vast range of materials.
Qualitest offers a wide variety of laboratory FTIR spectrometers for verify the identity and specifications of raw materials and products. The QualiFT-IR 5000 Spectrophotometer is built on an upgraded integration electronic system that is designed with highest flexibility and stability. It also comes with more powerful software, especially FDA CFR 21 program.
Features
- Powerful MainFTOS Suite software comes with specialized IR libraries, real-time instrument status diagnosis, modular platform design, FDA CFR 21 Part 11 Compliance, and auto software updating. It is also compatible with all windows versions.
- High sensitivity optical system ensuring long term stability, without the need of dynamic alignment which needs extra complicated electronic circuits.
- High stability modular partition design greatly improves the mechanical stability and long term working stability of the instrument.
- Intelligent multi-sealed moisture-proof design for preventing influences of high temperature, high humidity and chemical corrosions to the optical system in many ways.
- Innovated integration electronic system ensuring high quality real-time data collection and high-speed transmission.
- The electronic system is designed to meet CE certification and electromagnetic compatibility requirements, minimizing electromagnetic radiation in design and technology.
- High intensity, long lifetime IR source module, with the highest energy distributed in fingerprint region, adopts a reflex sphere design to obtain even and stable IR radiation.
- The sampling compartment reserving accessory mounting holes on the base plate, is designed for various accessories such as Defused/Specular Reflection, ATR, Liquid cell, GaS cell, and IR microscope etc.
Technical Specifications
Interferometer
|
Cube-corner Michelson interferometer
|
Beam splitter
|
Multilayer Gecoated KBr
|
Detector
|
Room temperature DLATGS module (standard)
|
Temperature stabilized, high sensitivity DLATGS module (optional)
|
IR Source
|
High intensity, long lifetime, air-cooled IR source
|
Wavenumber Range
|
7800 cm-1 ~ 350 cm-1
|
Resolution
|
0.85 cm-1
|
Signal to noise ratio
|
Better than 20,000:1 (RMS value, at 2100 cm-1 ~ 2000 cm-1 or 2100 cm-1 ~ 2200 cm-1, resolution: 4 cm-1, detector: DLATGS, 1 minute data collection)
|
Wavenumber Accuracy
|
±0.01 cm-1
|
Scanning Speed
|
Microprocessor control, different scanning speed selectable.
|
Software
|
MainFTOS Suite software workstation, compatible to all version Windows OS
|
Various specialized functional software modules (optional)
|
Communication
|
Ethernet interface (standard)
|
WIFI wireless communication (optional)
|
Data Output
|
Standard data format, report generation and output
|
Status Diagnosis
|
Power on self-check, real-time temperature and humidity monitoring and reminders
|
Certification
|
FDA 21 CFR Part11 compliance, IQ/OQ/PQ
|
Environment Conditions
|
Temperature: 10 ℃~30 ℃,
humidity: less than 70%
|
Dimensions & Weight
|
520 x 435 x 255 mm, 24 kg
|
Accessories
|
Transmission sample holder (Standard)
|
Optional accessories such as gas cell, liquid cell, Defused/Specular Reflection, single/multi reflection ATR, etc.
|