The X-CLARITY™ Tissue Clearing System II is an all-in-one, easy-to-use solution for electrophoretic tissue clearing system. Its unique design accelerates the removal of lipids from tissues while preserving the structural integrity of the sample.
Users can set tissue clearing conditions through a simple and intuitive touchscreen interface. In ETC (electrophoretic tissue clearing) mode, platinum-plated electrodes generate an electric field to accelerate the removal of lipids from tissues in a highly efficient manner.
A built-in temperature control system actively cools and heats buffer to maintain consistent buffer temperatures during clearing. Buffer is constantly circulated to ensure consistent buffering capacity, temperature control, and elimination of tissue clearing byproducts. This advanced system ensures efficient, rapid, and consistent tissue clearing.
Features
Precise temperature control
- Active buffer cooling and heating capacity
- Sensitive and accurate temperature sensor
|
Uniform electric field
- Platinum-plated electrodes
- Constant current and constant voltage modes
|
Compatible with multiple tissue types and sizes
- Electrophoretic and passive clearing
- Holders of various sizes available
|
User-friendly setup
- Simple touchscreen interface
- Ready-to-use clearing solution
|
Compatible sample holders
X-CLARITY™ ETC Chamber Specifications
Instrument Type |
Electrophoretic chamber |
External dimensions (W x D x H) |
176 mm x 128 mm x 154 mm |
Inernal dimensions (W x D x H) |
57 mm x 30 mm x 93 mm |
Weight |
2.8 kg |
X-CLARITY™ ETC Control Tower Specifications
Instrument Type |
Power supply, temperature control, and buffer circulation device |
User interface |
5 inch TFT LCD touchscreen |
Power supply modes |
Constant current or constant voltage |
Current & voltage range |
0.2-1.5 A, 5-70 V |
Temperature range |
30-60° C |
Pump speed range |
50-200 rpm |
Electrical requirements |
AC 100-240 V, 50/60 Hz |
Power consumption |
500 W (including the X-CLARITY™ ETC Chamber) |
Dimensions (W x D x H) |
205 mm x 430 mm x 370 mm |
Weight |
20 kg |