FDA recalls Huber infusion set needles

The U.S. Food and Drug Administration today announced a Class I recall of Exel/Exelint Huber needles, Exel/Exelint Huber Infusion Sets and Exel/Exelint "Securetouch+" Safety Huber Infusion Sets, manufactured by Nipro Medical Corporation for Exelint International Corporation.finf

Huber needles are used to access ports implanted under the skin of chronically ill patients for repeated access to veins for the withdrawal of blood and infusion of medication, nutritional solutions, blood products, and imaging solutions. These needles should be designed to penetrate the port without cutting and dislodging any silicone cores (or slivers) from the ports into which they are inserted.

Inspections conducted in October 2009 of Nipro facilities in Japan found that their needles "cored" in 60 to 72 percent of tests. The reason for this coring is related to design and manufacturing processes, which the FDA continues to investigate.

There are more than 2 million units impacted by this recall in distribution nationwide. Recalled needles were manufactured from January 2007 to August 2009. Units subject to recall have a lot number that begins with "07," "08," "09," and one of the following product codes or catalog numbers:

Exel/Exelint Huber Needles Product Codes/Catalog Numbers
   26901
   26902
   26904
   26906
   26907
   26908
   26909
   26911
   26921
   26922
   26923
   26924
   26925

Exel/Exelint Huber Infusion Sets Product Codes/Catalog Numbers
With/Injection Site:
   27940R
   27941R
   27944R
   27945R
   27946R
   27948R
   27949R
   27950R

Without/Injection Site Product Codes/Catalog Numbers:
   27954R
   27955R
   27958R
   27959R

Exel/Exelint "SecureTouch +" Safety Huber Sets Product Codes/Catalog Numbers:
   37854S
   37855S
   37858S

Hospitals, clinics and patients who have needles from these lists should immediately stop using these affected products and return any unused products to Exelint International Corporation. Direct all questions concerning this recall to:

Attn: Armand Hamid
EXEL International
5408 West Centinela Ave
Los Angeles, Calif. 90045-1504
Tel. 800-940-3935
Fax 800-308-5048
E-mail: [email protected]

Following hospital reports to the FDA of leakage after accessing the port with a Huber needle (labeled to be non-coring), the agency conducted it own laboratory testing of Huber needles from multiple manufacturers. This testing showed that certain Huber needles produced cores when inserted into ports. However, at this time only needles manufactured by Nipro have shown a high frequency for coring. The agency is continuing its investigation and will update the public if there are new developments.

At this time, the FDA has not received any adverse event reports related to silicone foreign bodies released in patients from Huber needle coring.The agency received only reports of port leakage. However, because it may be very difficult for clinicians to associate adverse patient outcomes with the use of defective Huber needles, there may be under-reporting of events.

The agency has issued a letter to manufacturers of other Huber needles to address design and manufacturing concerns.

"The agency's laboratory work helped determine the coring problem, and we will continue to work with manufacturers to address coring issues as quickly and thoroughly as possible," said Jeffrey Shuren, M.D., J.D., director of the FDA's Center for Devices and Radiological Health.

The FDA continues to work closely with all 20 manufacturers of Huber needles to understand the potential causes for coring and identify corrective measures. While a more thorough investigation of this situation continues, the FDA recommends that health care professionals consider taking the following precautions, which will be posted on the FDA's Web site, when accessing implanted ports with Huber needles:

  • Avoid flushing the syringe when initially confirming needles patency upon accessing the port. If the needle has cored the port, flushing may introduce the core into the patient's body, and could lead to serious adverse events.
  • When possible, upon accessing the port, consider aspirating a small amount of blood from the port after septum puncture, then discard the syringe with its contents. This step may recapture the silicone sliver. If the needle becomes clogged when attempting to aspirate, remove the needle, discard it and select a new one.
  •  Watch for signs and/or symptoms that may indicate damage to the port's septum, such as medication leakage resulting in inadequate therapy delivery, along with tissue, nerve and/or muscle damage; and redness of the surrounding area.
  • Patients should continue to follow their doctor's recommendations for receiving treatment.

Health care professionals and consumers may report serious adverse events (side effects) or product quality problems with the use of these products to the FDA's MedWatch Adverse Event Reporting program either online, by regular mail, fax or phone.

Comments

The opinions expressed here are the views of the writer and do not necessarily reflect the views and opinions of News Medical.
Post a new comment
Post

While we only use edited and approved content for Azthena answers, it may on occasions provide incorrect responses. Please confirm any data provided with the related suppliers or authors. We do not provide medical advice, if you search for medical information you must always consult a medical professional before acting on any information provided.

Your questions, but not your email details will be shared with OpenAI and retained for 30 days in accordance with their privacy principles.

Please do not ask questions that use sensitive or confidential information.

Read the full Terms & Conditions.

You might also like...
Cellular Origins, Fresenius Kabi sign development agreement for scalable automation of cell and gene therapy manufacturing