Multiple users in a Class II Biosafety Cabinet

This is an age old question that comes every now and then. First in reference from the BMBL (Biosafety in Microbiological and Biomedical Laboratories), a clear 12-inches (305 mm) should be present for separation of materials whether clean to dirty or between two users. “Therefore, as a general rule of thumb, keeping clean materials at least one foot away from aerosol-generating activities will minimize the potential for cross-contamination.” Second, if the work contains any volatiles or vapors, these would present exposure between the two operators as they are not retained in the HEPA filters. Lastly, use of proper aseptic technique is always important for the highest level of both product and personnel protection.

Multiple users in a Class II Biosafety Cabinet

Image credit: NuAire

A few additional comments, most E, H & S personnel would not advocate for two technicians work in the same cabinet. This perspective is because the two technicians most likely would be doing different research work and would be too difficult to control work practices, etc. to assure complete separation. If work is done with two technicians in a 6-foot Biosafety Cabinet (BSC), the work should be similar and work practices need to be adhered too!

Source:

Citations

Please use one of the following formats to cite this article in your essay, paper or report:

  • APA

    NuAire. (2020, May 12). Multiple users in a Class II Biosafety Cabinet. News-Medical. Retrieved on November 21, 2024 from https://www.news-medical.net/news/20190820/Multiple-users-in-a-Class-II-Biosafety-Cabinet.aspx.

  • MLA

    NuAire. "Multiple users in a Class II Biosafety Cabinet". News-Medical. 21 November 2024. <https://www.news-medical.net/news/20190820/Multiple-users-in-a-Class-II-Biosafety-Cabinet.aspx>.

  • Chicago

    NuAire. "Multiple users in a Class II Biosafety Cabinet". News-Medical. https://www.news-medical.net/news/20190820/Multiple-users-in-a-Class-II-Biosafety-Cabinet.aspx. (accessed November 21, 2024).

  • Harvard

    NuAire. 2020. Multiple users in a Class II Biosafety Cabinet. News-Medical, viewed 21 November 2024, https://www.news-medical.net/news/20190820/Multiple-users-in-a-Class-II-Biosafety-Cabinet.aspx.

Comments

The opinions expressed here are the views of the writer and do not necessarily reflect the views and opinions of News Medical.
Post a new comment
Post

While we only use edited and approved content for Azthena answers, it may on occasions provide incorrect responses. Please confirm any data provided with the related suppliers or authors. We do not provide medical advice, if you search for medical information you must always consult a medical professional before acting on any information provided.

Your questions, but not your email details will be shared with OpenAI and retained for 30 days in accordance with their privacy principles.

Please do not ask questions that use sensitive or confidential information.

Read the full Terms & Conditions.

You might also like...
NuAire now a supplier of E&I Cooperative Services