Achieve the fastest separations possible in the shortest amount of time with the Avanti® J-30I high performance centrifuge

  • Samples spend more time at full force and less time in the centrifuge
  • Swinging-bucket and fixed-angle rotors provide maximum separation forces in excess of 100,000 x g at speeds up to 30,000 rpm
  • Unmatched acceleration/deceleration rates
  • Easy run set-up

Features

Accelerate your discovery with the fastest separations possible.

Near-Ultracentrifuge Forces

  • Quickly and easily process protein separations with over 100,000 x g and JA-30.50, highest g-force fixed-angle rotor available in a high-performance centrifuge
  • Separate subcellular organelles with rate zonal centrifugation and the JS-24.38, a 6 x 38 mL swinging-bucket rotor

High-Throughput Processing

  • Four-liter batch throughput for bacteria and cell membrane isolation using the JLA-9.1000 J-LITE rotor at 16,800 x g
  • DNA sample prep in up-to ten microplates with the JS-5.9 rotor

Fast and Efficient Design

  • Fastest acceleration and deceleration times with exclusive Switched Reluctance Drive** technology
  • Low-heat output and low-energy consumption

Avanti J-30I Centrifuge System Specifications

UNSPSC

41103905

Height

86 cm (34 “); with door open 141 cm (55.5”)

Width

71 cm (28”)

Depth

86 cm (34”); 102 cm (40.25”) including air diverter extending from back panel

Weight

310 kg (680 lb)

Friction Reduction System (FRS)

310 kg (680 lb)

Approximate acceleration time

Maximum, slow, or timed (1-10 min from 0 to 500 rpm)

Approximate deceleration time

Maximum, slow, timed (1-10 min from 500 to 0 rpm), or off

Noise level (0.91meters /3' in front of centrifuge)

57 dBA

Max heat dissipation into room (steady-state conditions)

2.0 kW (6,900 Btu/hour)

Set time

1 min-99 hour 59 min, or continuous (hold)

Set temperature

-20 to +40°C in 1°C increments

Set speed

Actual rotor speed ± 10 rpm from 10,000 to 30,000 rpm or equivalent RCF

Temperature control

± 2°C of set temperature (after equilibration)

Ambient temperature range

16 to 38°C

Clearances required

7.6 cm (3.0 in) sides; 16.0 cm (6.25 in) rear

Door

5.4-cm (2.1-in) thick structural foam with steel plate

Humidity restrictions

< 95% (noncondensing)

Pollution degree

2

Rotor chamber diameter

40.6 cm (16”)

Time–actual display

Indicates run time remaining (timed run) or elapsed time (hold run)

Other Equipment by this Supplier

While we only use edited and approved content for Azthena answers, it may on occasions provide incorrect responses. Please confirm any data provided with the related suppliers or authors. We do not provide medical advice, if you search for medical information you must always consult a medical professional before acting on any information provided.

Your questions, but not your email details will be shared with OpenAI and retained for 30 days in accordance with their privacy principles.

Please do not ask questions that use sensitive or confidential information.

Read the full Terms & Conditions.