The DNBSEQ-T20×2 is designed to accommodate the most difficult sequencing circumstances. Using OMICS technology, global large-scale population genome studies and healthcare can be improved. The fully automated sequencer DNBSEQ-T20×2 processes six slides per run using robotic integration of all genetic sequencer components.

It addresses all high-throughput sequencing requirements in both academic and medical applications, including those for WGS, WGBS, stLFR, single-cell sequencing, stereo-seq, etc.

Features

DNBSEQ-T20×2: Fully automated sequencer for genome projects

Image Credit: MGI

  • Six slides run simultaneously
  • >50,000 WGS/year
  • 30× Genome coverage

Performance parameter

Table 1. Source: MGI

DNBSEQ-T20×2RS
Model DNBSEQ-T20×2RS
Slide/ Run 6
Effective Reads/Slide* 35 B (PE100)
40 B (PE150)
Read lengths PE100
PE150

 

* The maximum number of effective reads are based on a specific standard library, and the actual application library will fluctuate depending on the sample type and library construction method.

Table 2. Source: MGI

DNBSEQ-T20×2RS
Supported Read Lengths Data Output per run Data Quality Q30* Run Time per run**
PE100 ~42 Tb >85% ~60 h
PE150*** ~72 Tb >80% ~80 h

 

* The percentage of bases above Q30 is obtained by averaging metrics for runs using standard library. Actual performance is influenced by sample type, library quality and type, insert length among other factors.

** The run time includes DNB loading, sequencing, and generation of Cal. file. Cal. file is a binary file format generated by MGI sequencer basecall software.

*** PE150 will be released on Q3 2023.

Product components

DNBSEQ-T20×2: Fully automated sequencer for genome projects

Image Credit: MGI

System parameter

Table 3. Source: MGI

Power, Dimensions & Net Weight
Power Type: 380/400 V, 50Hz
Rated Power: 20,000 VA
Dimension: 4200 mm (L)* 4800 mm (W) * 2000 mm (H)
Net Weight: 2743 kg (without BIT module)
Floor Load Capacity: 500 kg/m2

 

Table 4. Source: MGI

Operating Environment Requirements
Temperature: 19~22 °C
Relative Humidity: 30%RH ~ 70%RH
Atmospheric Pressure: 90 kPa~106 kPa
Pollution Level: Level 2

 

*Can only be used indoors.

 


*Unless otherwise informed, StandardMPS and CoolMPS sequencing reagents, and sequencers for use with such reagents, are not available in Germany, Spain, UK, Sweden, Belgium, Italy, Finland, Czech Republic, Switzerland, Portugal, Austria, and Romania. Unless otherwise informed, StandardMPS sequencing reagents, and sequencers for use with such reagents are not available in Hong Kong.

*Products are provided for Research Use Only. Not for use in diagnostic procedures (except as specifically noted).

Other Equipment by this Supplier

While we only use edited and approved content for Azthena answers, it may on occasions provide incorrect responses. Please confirm any data provided with the related suppliers or authors. We do not provide medical advice, if you search for medical information you must always consult a medical professional before acting on any information provided.

Your questions, but not your email details will be shared with OpenAI and retained for 30 days in accordance with their privacy principles.

Please do not ask questions that use sensitive or confidential information.

Read the full Terms & Conditions.