MGISTP-3000 from MGI is considered a benchtop automated sample transfer processing system. It combines tube decapping, barcode identification, tube recapping, automated liquid transfer, and negative pressure protection.

MGISTP-3000 provides high torque decapping gripper, 23-in touch screen, and HEPA filters, ensuring the utmost filtration efficiency of 99.995% at 0.3 µm. Moreover, MGISTP-3000 enables users to load three types of tubes in the same run.

Users can transfer 96 samples from screw cap transport tubes to 96-Well Microplates in nearly 40 minutes. A HEPA filter purifies inflow and exhaust air to avoid aerosol pollution and ensure a secure working environment for nucleic acid extraction and laboratory technicians.

Features

Automated: Completely automated—offers tube decapping, tube recapping, barcode identification, and automated liquid transfer functions.

Easy to use: Comes with a user-friendly GUI touch screen and enables users to load three types of tubes in a similar run.

Safe: Two HEPA-filtration and one UV light guarantee a more secure working environment for both liquid transfer and laboratory technicians.

High Reliability: Users can automatically log common errors without requiring human intervention to guarantee continuous functioning.

MGISTP-3000: Benchtop automated sample transfer processing system

Image Credit: MGI

Performance parameter

Table 1. Source: MGI

MGISTP-3000
Throughput 96 Samples/40 min
Pipetting Method Based on the principle of air displacement
Volume Range 10 μL:<±10% ; 100 μL:<±5%; 1000 μL:<±2%
Accuracy 10 μL:<±10% ; 100 μL:<±5%; 1000 μL:<±2%
CV 10 μL:<5% ; 100 μL:<2%; 1000 μL:<1%
Robot Arm Accuracy X-Y-Z:±0.1 mm
Barcode Scanner 1D and 2D
UV Light 3*8 W
Supply/Exhaust Air Filter Exhaust / Inflow HEPA filtration efficiency of 99.995% at 0.3um, Negative Pressure System
Tube Compatibility 5-15 mL transport tube with screw cap.
Functions Integrate multiple functions, including decapping, recapping, barcode identification, automated liquid transfer, HEPA-filtration and negative pressure protection.

 

System parameter

Table 2. Source: MGI

. .
Power & Dimensions & Net Weight Dimensions: Length 1100 mm, Width 680 mm, Height 1050 mm
Net Weight: <200 kg
Power Type: 100 ~ 240 V,50/60 Hz
Rated Power: 600 VA
Operating Environment Requirements Temperature: 19-25 ℃
Relative Humidity: 20% RH ~ 80% RH, non-condensing
Atmospheric Pressure: 80 ~ 106 kPa

Other Equipment by this Supplier

While we only use edited and approved content for Azthena answers, it may on occasions provide incorrect responses. Please confirm any data provided with the related suppliers or authors. We do not provide medical advice, if you search for medical information you must always consult a medical professional before acting on any information provided.