Patients likely to experience inappropriate shocks from defibrillators without firmware update

A case study of a patient who experienced inappropriate shocks from her defibrillator is presented in Heart Rhythm Case Reports, an official journal of the Heart Rhythm Society, published by Elsevier. This event likely took place because an FDA-recommended firmware update to strengthen cybersecurity had not occurred. This underscores the importance of upgrading firmware of Abbott devices according to FDA recommendations.

The patient, with an Abbott Fortify AssuraTM implantable cardioverter-defibrillator (ICD) with a Merlin@homeTM radiofrequency communicator, presented to the emergency department after receiving two shocks from her ICD without preceding symptoms. She had a history of atrial fibrillation with rapid ventricular response.

The patient was enrolled in at-home remote monitoring for her device and had frequent in-person device checks; however, the patient's device had outdated ICD firmware.

In August 2016, Muddy Waters LLC, an investment firm that conducts investigative research on public companies, released a report claiming that certain St. Jude Medical/Abbott cardiovascular implantable electronic devices (CIEDs) were vulnerable to cyberattack through the Merlin@home™ radiofrequency remote monitoring system, which allows care teams to review medical and technical information about the patient and the device without an in-person visit. Senior investigator Vineet Kumar, MD, FHRS, Division of Cardiac Electrophysiology, Inova Heart and Vascular Institute, Falls Church, VA, USA, explained, "Cyberattack of CIEDs could affect patients' confidentiality, interrupt remote monitoring, and even harm patients by changing device settings or promoting early battery depletion."

Consequently, St. Jude Medical/Abbott released a software patch for the radiofrequency communicator, which was successfully programmed remotely into nearly 100% of actively used Merlin@home™ radiofrequency communicators. The company later released firmware updates to strengthen cybersecurity performance in the devices themselves. This requires an in-person visit to the healthcare provider, but it takes only three minutes to complete and is rarely associated with complications. Still, the firmware has only been updated in 24% of eligible devices. Because no harm is known to have been caused by a CIED cyberattack, deferring the firmware update may not have been prioritized for many patients. Additionally, reports have emerged showing that the firmware update may cause irreversible device malfunction with an incidence of 0.003%.

When the patient arrived at the emergency department, she was asymptomatic, and her vital signs were normal. Device interrogation demonstrated the ICD programming had reverted to backup mode, and thus no electrocardiograms were recorded during her shocks. Her device had several radiofrequency connection/disconnection events with the Merlin@home™ system over a short period of time. This was detected as a potential cyberattack and the device entered backup mode to avoid cybersecurity vulnerabilities due to event queue overload (EQO). In backup mode, the device is automatically reprogrammed to treat any heart rhythm with a rate over 146 beats per minute (BPM) with a shock. Based on the patient's history of atrial fibrillation with rates over 150 BPM, inappropriate treatment of atrial fibrillation with rapid ventricular response is the most likely cause of her ICD shocks.

EQO events occur most frequently in the setting of an updated Merlin@homeTM software patch being used with outdated ICD firmware. This combination is currently in use in almost 75% of affected Abbott ICDs.

The patient's ICD was reprogrammed to the original settings, the cybersecurity firmware was upgraded, and she was released from the emergency department.

Physicians and their patients with affected Abbott devices now have another reason to consider updating their device firmware. While this is the first reported case, based on the reported frequencies of EQO events and the frequency of outdated firmware still in use in affected devices, other patients may experience similar events. This case highlights the importance of following FDA recommendations to update CIED firmware to protect not only against a cyberattack, but potentially even more importantly, to avoid unnecessary right ventricular pacing and ICD shocks."

Brett Atwater, MD, Co-Investigator, Director of Electrophysiology, Inova Heart and Vascular Institute in Falls Church, VA, USA

The investigators recommend that the possibility of an inappropriate shock and/or unnecessary right ventricular pacing be incorporated into patient discussions about the risks and benefits of firmware update, to better assist shared decision making.

Source:
Journal reference:

Qian, X., et al. (2022) Radiofrequency remote monitor software patch update without cybersecurity implantable cardioverter-defibrillator firmware update increases the risk of inappropriate implantable cardioverter-defibrillator therapies. Heart Rhythm Case Reports. doi.org/10.1016/j.hrcr.2021.12.016.

Comments

The opinions expressed here are the views of the writer and do not necessarily reflect the views and opinions of News Medical.
Post a new comment
Post

While we only use edited and approved content for Azthena answers, it may on occasions provide incorrect responses. Please confirm any data provided with the related suppliers or authors. We do not provide medical advice, if you search for medical information you must always consult a medical professional before acting on any information provided.

Your questions, but not your email details will be shared with OpenAI and retained for 30 days in accordance with their privacy principles.

Please do not ask questions that use sensitive or confidential information.

Read the full Terms & Conditions.

You might also like...
New research supports more liberal blood transfusion approach for heart attack patients