NIST: The Model Antibody for Glycoform Profiling

There are many different scenarios in biopharmaceutical drug discovery and development that need fast, accurate and reliable analysis of antibodies by mass spectrometry, from innovator and biosimilar development to quality control of production.

BioPharma Compass software platform, which emphasizes the simplification and streamlining of automated data acquisition and processing for common analytical workflows, was developed in response to researcher expectations for rapid results with the lowest possible interaction effort.

To demonstrate the pace at which a sizeable quantity of valuable data can be produced and interpreted in BPC 2.0, the NIST reference antibody (1) was measured with the Bruker maXis ultra-high resolution QTOF mass spectrometer by means of a trio of complementary approaches that offered unique abilities for providing detailed information at great speed (Fig. 1).

Graphical abstract, showing the use of BioPharma Compass 2.0 for automated antibody characterization using intact, subunit domain and bottom-up approaches. The software also contains other features not shown here, including workflows for MALDI and ETD, and a Regulatory Toolkit.

Figure 1. Graphical abstract, showing the use of BioPharma Compass 2.0 for automated antibody characterization using intact, subunit domain and bottom-up approaches. The software also contains other features not shown here, including workflows for MALDI and ETD, and a Regulatory Toolkit.

Results and Discussion

Intact Mass Screening

Valuable data relating to the major glycoforms present in the sample was produced through analysis of the intact NIST antibody by UHPLC-MS, with just a simple 1 minute desalting LC gradient (total run time 10 minutes).

For example, the key glycoforms G0F, G1F and G2F were marked up, lysine clipping could easily be identified for each glycoform, and even the aglycon was detected, signifying a good dynamic range of the analysis.

Each of these signals was routinely detected with a mass accuracy of < 2 Da, offering a confident impression of the quality and composition of the sample over a brief time frame (Fig. 2).

Intact mass screening of NIST antibody. Maximum entropy deconvoluted spectrum shows glycoforms detected across /> 2 orders of magnitude. Insert shows low abundant aglycon forms in more detail.

Figure 2. Intact mass screening of NIST antibody. Maximum entropy deconvoluted spectrum shows glycoforms detected across > 2 orders of magnitude. Insert shows low abundant aglycon forms in more detail.

Experimental Details  
Platform: BPC 2.0 software, maXis II UHR-QTOF, Standard flow UHPLC, BEH C4 2.1 x 100 mm 1.7 µm (Waters) or C18+ 2.1 x 100 mm 1.5 µm (Thermo) column (see below), Solvent A 0.1% FA in water, Solvent B 0.1% FA in acetonitrile.
Intact mass: 1 µg injected on C4 column heated to 80 °C, 0.5 mL/min, 27-95% B gradient in 1 min, total run time 10 mins.
Middle-Down: Digestion with IdeS (FabRICATOR, Genovis) , followed by DTT reduction in guanidine-HCl. 1 µg injected on C4 column heated to 40 °C, 0.2 mL/min, 27-42% B gradient in 20 min after equilibration to waste for 20 min at 2% B to remove guanidine-HCl.
Bottom-Up: Digestion with Trypsin. 4 µg injected on C18 column heated to 40 °C, 0.2 mL/min, 2-40% B in 45 min, MS 2 Hz, MS/MS 2-5 Hz depending on precursor intensity.

 

Subunit Domains

In the past, a domain-based approach has been used to study antibody intact mass inconsistencies stemming from sequence variants (2).

In this instance, it is employed for simple confirmation and relative quantitation of the modifications existent on the respective antibody subunit, for example, glycation on the LC and pyroglutamylation on the Fd subunit (Fig. 3).

Subunit domain analysis of NIST antibody. Left from top, LC-MS TIC chromatogram, NIST LC acquired isotopic pattern (blue) overlaid with theoretical (red), table of results (abridged). Right from top, maximum entropy deconvoluted spectra for NIST Fc/2 (insert shows relative quantitation of lysine clipping), LC and Fd subunits.

Figure 3. Subunit domain analysis of NIST antibody. Left from top, LC-MS TIC chromatogram, NIST LC acquired isotopic pattern (blue) overlaid with theoretical (red), table of results (abridged). Right from top, maximum entropy deconvoluted spectra for NIST Fc/2 (insert shows relative quantitation of lysine clipping), LC and Fd subunits.

Interestingly, the inlay in Fig. 3 illustrates that 91.2 % of the mAb, typified by the G1F glycoform, had experienced lysine clipping.

The major benefit of the subunit method is that Bruker’s SNAP algorithm determines monoisotopic molecular weights (in the majority of cases with sub ppm mass accuracy) from isotopic peaks with high isotopic fidelity (True Isotopic Pattern, or TIP™) that were resolved to the baseline (Table in Fig 3).

This allows for protein modifications to be quantified with ease and accuracy, and even facilitates the investigation of deamidation events on the domain level. Both anticipated and unanticipated protein modifications can be confidently assigned in BPC 2.0, and can be corroborated further by overlaying the theoretical isotope pattern.

Bottom-Up Analysis

Sequence validation of antibodies calls for the fragmentation of all available peptides, so as to attain near 100 % sequence coverage. The fragment spectra generated must also be of exceptional quality in order to deliver fragment peak evidence (fragment coverage) for every individual amino acid.

The NIST antibody was digested solely with trypsin, and the peptides were separated with a comparatively brief 45 minute gradient. Fragmentation was activated with an Instant Expertise™ auto MS/MS acquisition method and produced 99.5 % sequence coverage for the light chain (LC) in one experiment (Fig. 4).

Bottom up analysis of NIST antibody showing near 100% sequence coverage. From top, example MS/MS spectrum, sequence coverage map and an excerpt from the BPC 2.0 results table

Figure 4. Bottom up analysis of NIST antibody showing near 100% sequence coverage. From top, example MS/MS spectrum, sequence coverage map and an excerpt from the BPC 2.0 results table

This outcome was achieved with MS/MS fragment spectra that displayed outstanding detection of low and high mass fragments at the same time, which led to 96.2% fragment coverage of the NIST LC.

Conclusion

BioPharma Compass 2.0 and the maXis II were employed for the automated evaluation of the NIST antibody, using a trio of complementary approaches: intact, subunit domain and bottom-up. This grouping delivered rapid, precise and detailed data for antibody analysis using just one software platform.

BPC 2.0 was designed to be extremely simple to operate, while upholding the experimental flexibility necessary in the modern biopharmaceutical analytical laboratory. The software also facilitates 21 CFR Part 11 compliant work.

The specific benefits of each of the described approaches are:

Intact mass: Rapid screening for anticipated glycoforms, including low abundant species such as the aglycan and prompt detection of antibody processing artifacts.

Subunit domain: Precise mass analysis of antibody subunits on the sub ppm level, glycoform quantitation and additional localization of sequence deviations on the Fd, Fc/2 or LC domains (2).

Bottom up: Close to 100% sequence and fragment coverage for sequence validation from just one experiment.

References

  1. Formolo T et al. (2015), Determination of the NISTmAb Primary Structure, in: State-of-the-Art and Emerging Technologies for Therapeutic Monoclonal Antibody Characterization Volume 2. Biopharmaceutical Characterization: The NISTmAb Case Study, Chapter 1, pp 1–62, ACS Symposium Series, Vol. 1201; Chapter DOI: 10.1021/bk-2015-1201.ch001.
  2. Ayoub D et al. Correct primary structure assessment and extensive glyco-profiling of cetuximab by a combination of intact, middle-up, middle-down and bottom-up ESI and MALDI mass spectrometry techniques. MAbs. 2013;5(5):699-710. doi: 10.4161/ mabs.25423.

Some of the work reported here was undertaken as part of a project for the Consortium for Top-Down Proteomics: “Structural Analysis of Monoclonal Antibodies by Top-Down Mass Spectrometry: An Intra-Laboratory Study”.

About Bruker Life Sciences Mass Spectrometry

Discover new ways to apply mass spectrometry to today’s most pressing analytical challenges. Innovations such as Trapped Ion Mobility (TIMS), smartbeam and scanning lasers for MALDI-MS Imaging that deliver true pixel fidelity, and eXtreme Resolution FTMS (XR) technology capable to reveal Isotopic Fine Structure (IFS) signatures are pushing scientific exploration to new heights. Bruker's mass spectrometry solutions enable scientists to make breakthrough discoveries and gain deeper insights.


Sponsored Content Policy: News-Medical.net publishes articles and related content that may be derived from sources where we have existing commercial relationships, provided such content adds value to the core editorial ethos of News-Medical.Net which is to educate and inform site visitors interested in medical research, science, medical devices and treatments.

Last updated: Feb 15, 2023 at 12:10 AM

Citations

Please use one of the following formats to cite this article in your essay, paper or report:

  • APA

    Bruker Life Sciences Mass Spectrometry. (2023, February 15). NIST: The Model Antibody for Glycoform Profiling. News-Medical. Retrieved on December 22, 2024 from https://www.news-medical.net/whitepaper/20190402/NIST-The-Model-Antibody-for-Glycoform-Profiling.aspx.

  • MLA

    Bruker Life Sciences Mass Spectrometry. "NIST: The Model Antibody for Glycoform Profiling". News-Medical. 22 December 2024. <https://www.news-medical.net/whitepaper/20190402/NIST-The-Model-Antibody-for-Glycoform-Profiling.aspx>.

  • Chicago

    Bruker Life Sciences Mass Spectrometry. "NIST: The Model Antibody for Glycoform Profiling". News-Medical. https://www.news-medical.net/whitepaper/20190402/NIST-The-Model-Antibody-for-Glycoform-Profiling.aspx. (accessed December 22, 2024).

  • Harvard

    Bruker Life Sciences Mass Spectrometry. 2023. NIST: The Model Antibody for Glycoform Profiling. News-Medical, viewed 22 December 2024, https://www.news-medical.net/whitepaper/20190402/NIST-The-Model-Antibody-for-Glycoform-Profiling.aspx.

Other White Papers by this Supplier

While we only use edited and approved content for Azthena answers, it may on occasions provide incorrect responses. Please confirm any data provided with the related suppliers or authors. We do not provide medical advice, if you search for medical information you must always consult a medical professional before acting on any information provided.

Your questions, but not your email details will be shared with OpenAI and retained for 30 days in accordance with their privacy principles.

Please do not ask questions that use sensitive or confidential information.

Read the full Terms & Conditions.